Hello! Welcome To Standard Group (Hong Kong) Ltd.
Automobile Interior Material Testing InstrumentsProfessional testing equipment manufacturer
Service Hotline:+86(21)-64208466
Current location: Home Page  > Product > Automotive Wire Tester

Automotive Wire Combustion Tester

The Automotive Wire Combustion Tester is designed to evaluate the flame resistance and burning performance of automotive wires and cables. It ensures compliance...

The Automotive Wire Combustion Tester is designed to evaluate the flame resistance and burning performance of automotive wires and cables. It ensures compliance with international standards and helps verify wire safety under fire exposure conditions.

Application

(1) Automotive wire harnesses – assess flame resistance for vehicle safety.
(2) Low-voltage electrical cables – verify compliance with combustion standards.
(3) R&D laboratories – evaluate new wire materials for fire safety.
(4) Production quality control – ensure each batch meets flammability requirements.
(5) Certification testing – for ISO, SAE, DIN, and QC/T compliance verification.

Standards

(1) QC/T703-2005 – Automotive electrical wire flame resistance standard
(2) SAE J1128-2005 – Automotive wire flammability test
(3) DIN 72551 – Wire and cable combustion test
(4) ISO 6722 – Road vehicle electrical cables flame resistance

Parameters

Parameter Specification
Test Chamber Enclosed; dimensions: 650 × 450 × 900 mm; stainless steel with matte black background
Fuel Gas High-purity propane or liquefied petroleum gas
Test Sample Length 600 mm (24 in)
Sample Clamp Angle 45°
Bunsen Burner Tube Nominal diameter 9 mm; outer flame: 100 mm (4 in); inner flame: 50 mm (1/2 in); inner flame temperature: 900 ±50°C
Burner Position Below the sample, perpendicular to sample axis
Ignition Automatic
Vertical Flame Duration 15 s
Temperature Measurement High-precision K-type thermocouple with TES-1310 temperature meter
Exhaust System Ventilation fan removes combustion gases after test
JIS3005 Horizontal Combustion Fixture Included; horizontal combustion test setup
Horizontal Test Chamber 305 × 355 × 610 mm
Horizontal Sample Length 300 mm
Horizontal Sample Placement Horizontal
Horizontal Flame Duration 30 s
Horizontal Burner Tube Diameter 10 mm; inner flame: 35 mm; outer flame: 130 mm
Stop Method Automatic when preset data reached