Solution configuration method for salt spray test

Solution configuration method for salt spray test

The salt spray test chamber (LRHS-108-RY) can be used for neutral salt spray, acidic salt spray, and copper accelerated vinegar salt spray tests. As for the preparation of salt solution, the editor will prepare it for you: Regarding the preparation of neutral salt solution in the salt spray test chamber

① Dissolve analytical pure sodium chloride in pure water at a concentration of (50 ± 10) g/L;

② Measure the pH value of the test solution using a pH test strip at 25 ℃, ranging from 6.5 to 7.5;

③ When it exceeds the range, analytical pure hydrochloric acid or sodium hydroxide solution can be added for adjustment. The prepared solution must be filtered before use.

Regarding the preparation of acetate solution in the salt spray test chamber

① Firstly, dissolve the analytical pure sodium chloride in the pure water of the weathering test equipment, with a concentration of (50 ± 10) g/L;

② Then adjust the pH value of the solution with acetic acid to be within the range of 3.1 to 3.3, which is measured using the collected salt spray sample;

③ The pH value should be measured at 25 ℃, and the prepared solution must be filtered before use.