The Mask Integrity Test Accessory (MITA) works in tandem with the CBRN Mask Protection Assessment Test System or PortaCount? Respirator Fit Tester to quickly detect leaks in CBRN gas masks and commercial respirators and ensure your personnel are getting maximum protection.
MITA assesses the overall integrity of masks by performing an aerosol leak test utilizing the PortaCount Respirator Fit Tester as a particle detector. While under conditions of negative pressure and constant flow, the mask is tested on a full headform with an enhanced bladder seal system to provide a tight seal for most CBRN gas mask and commercial respirator types and sizes. Testing masks on MITA’s full headform configuration allows for use of the mask’s head harness, mimicking a real-life application and providing more accurate test results.
If a leak is detected during the assessment, the integrated aerosol wand and the generator are used to pinpoint invisible leaks for quick evaluation and repair. The operator uses the aerosol wand to direct a stream of concentrated particles at specific locations around the mask until the single-particle counting capability of the PortaCount Respirator Fit Tester detects the suddenly elevated particle concentration and alerts the operator.
APPLICATIONS:
Mask integrity testing
Exhalation valve testing
Drink tube testing
INCLUDED ITEMS:
Universal AC Adapter with Line-Cord
Headform
Headform Mask Sample Plug
#1 and #3 Rubber Stoppers
Mask Bypass Hose
Twin Tube with 1 in. relief cut
Aerosol Wand
Aerosol Generator Assembly
Aerosol Containment Hood
Aerosol Mixer
Mask Filter Plug
HEPA Filter Canister (x2)
Drink Tube Sampling Hose
DT & EV Bypass Hose
Emery Oil (500 ml)
PortaCount Fit Tester Interface Cable (for Models 8030/8038 only)
MITA Computer Cable
USB to Serial Converter
MITA Communicator Software CD
Quick Start Guide
Self Check Components Reference Card
Operators' Manual
Calibration Certificate
One-Year Warranty
FEATURES & BENEFITS:
Separate tests for overall mask integrity (Protection Factor), exhalation valve and drink tube
Integrated aerosol wand pinpoints leak locations, allowing for immediate evaluation and repair
Full headform configuration allows for use of the mask’s head harness, mimicking real-life application and providing more accurate test results.
Single head form with an enhanced bladder seal system creates a tight seal for most gas mask and full face respirator types and sizes.
Easy-to-read display with an intuitive step-by-step test sequence allows for minimal operator training
Streamlined menu structure and no-nonsense keypad eliminate redundant button pushing
Quick setup/teardown process saves time
Efficient test sequence allows for 12+ mask assessments per hour
Packaged for ease of transport with one carrying case for the instrument and all components