Static Honest Meter is a measuring instrument for attenuation of static electricity, most appropriate for investigating the diffusibility of static.
This Static Honest Meter is used to electrify the specimen by irradiating it by air ions generated by corona discharges initiated by the device, and then, after the irradiation is stopped, it is used to investigate the decay curve of the charge on the specimen.
Surface Resistivity and Static Decay tester is an analyzing and computing unit which can automatically compute the half-life period of voltage decay from the static decay carve obtained from the host meter.
By connecting the analyzer and the honest meter together, it is possible to measure the attenuation of materials automatically in one-touch action.
Optional data analysis software can provide more detailed information and results.
﹣Applied on JIS L1049-1980 Electrostatic Propensity Testing Method for woven fabric and knitted work.
﹣Compared with the friction electrification or direct voltage applying electrification methods, the corona discharge method has excellent reproducibility, making it possible to compare the specimen with those of other methods.
﹣All specimens in sheet form, such as films, board, thread, cloth, carpet, or glass panes, can be subjected to measurement by using this device.
﹣By selecting the polarity and applied voltage for corona discharge, it is possible to electrify the specimen with either positive or negative charge. It is also possible to change the voltage value.
﹣Without any friction, no damages will be caused to the specimen, and accurate and repeated measurement is possible.
Standards:
GB/T 12703.1
FZ/T 01042
JIS L1094
Technical Parameters:
Items |
Parameters |
The measuring range of electrostatic voltage |
0 ~ 10kV |
The half-life time range |
0 ~ 99.99 h |
Rotation speed |
1500 RPM |
Discharge time |
0 ~ 99.99 s (test is scheduled (30 + 0.1) seconds) |
Discharge distance between needle electrode and specimen |
20mm |
Measurement distance between test probe and specimen |
15mm |
The static voltage calibration was performed before the instrument was issued
Check before use. The cylinder must be filled with oil (transformer oil) |
Sample size |
60mm x 80mm |
Power |
220V 50Hz 1.5kw
|