Principle and standard of synthetic blood penetration resistance tester
time:2021-02-02 10:10Browse:118Times
The Synthetic Blood Penetration Resistance Tester is one of the most popular products in Qinsun's textile testing equipment. Today, I will tell you some basic knowledge about the Synthetic Blood Penetration Tester, so that you can know how to choose a suitable equipment if you need it in the future.
The protective clothing material was tested with synthetic blood under continuously applied pressure, and the penetration of synthetic blood on the material was visually inspected.
Synthetic blood penetration tester for medical protective clothing is mainly used to test the resistance of protective clothing to synthetic blood penetration under different test pressures.The synthetic blood penetration tester for protective clothing is suitable for measuring the resistance of daily work clothes, surgical clothes, isolation gown and protective clothing to synthetic blood penetration under different test pressures.
Applicable standard of synthetic blood penetrability tester for protective clothing;
GB 19082-2009 technical requirements for medical disposable protective clothing-synthetic blood penetration test method
YY 0700-2008 Test Method of Synthetic Blood for Anti-penetration Performance of Protective Clothing Materials for Blood and Body Fluid Protective Equipment
ISO 16603:2004 blood and body fluid protective equipment protective clothing material resistance to blood and body fluid penetration test synthetic blood test method
ASTM F1670,ASTM F1671
Main parameters of Nonwovens Synthetic Blood Penetration Tester for protective clothing:
1. the air source which can provide (20 1) kpa air pressure is used to continuously pressurize the sample, which is not limited by the space of the test site;
2. It has a pressure gauge to display the pressurization pressure, and the pressurization pressure can be adjusted;
3. Use pressurized medium: compressed air;
4. The special stainless steel penetration test tank ensures that the sample is firmly clamped and prevents the synthetic blood from splashing around;
5. Square metal blocking net: open space ≥ 50%; Bending at 14kPa ≤ 5 mm;